HIV Research - Human Immunodeficiency Virus, Testing, Treatment, Prevention, Vaccines, AIDS

HIV Research Today is a free monthly online journal that collates and summarizes the latest research about HIV, including details on human immunodeficiency virus, testing, treatment, prevention, vaccines, aids.


HIV Research Today

Home

View Latest Issue

Information About HIV

Books on HIV

Advertising in Research Today

View Other Research Today Publications



Evaluation of the VACUTAINER PPT Plasma Preparation Tube for use with the Bayer VERSANT assay for quantification of human immunodeficiency virus type 1 RNA.

Elbeik T, Nassos P, Kipnis P, Haller B, Ng VL

Clinical Laboratories at San Francisco General Hospital, Department of Laboratory Medicine, Building NH, Room 2M35, 1001 Potrero Ave., San Francisco, CA 94110, USA. elbeik@itsa.ucsf.edu

Separation and storage of plasma within 2 h of phlebotomy is required for the VACUTAINER PPT Plasma Preparation Tube (PPT) versus 4 h for the predecessor VACUTAINER EDTA tube for human immunodeficiency virus type 1 (HIV-1) viral load (HIVL) testing by the VERSANT HIV-1 RNA 3.0 assay (branched DNA). The 2-h limit for PPT imposes time constraints for handling and transporting to the testing laboratory. This study compares HIVL reproducibility from matched blood in EDTA tubes and PPTs and between PPT pairs following processing within 4 h of phlebotomy, stability of plasma HIV-1 RNA at 24- and 72-h room temperature storage in the tube, and comparative labor and supply requirements. Blood from 159 patients was collected in paired tubes (EDTA/PPT or PPT/PPT): 86 paired EDTA tubes and PPTs were processed 4 h following phlebotomy and their HIVLs were compared, 42 paired PPT/PPT pairs were analyzed for intertube HIVL reproducibility, and 31 PPT/PPT pairs were analyzed for HIV-1 RNA stability by HIVL. Labor and supply requirements were compared between PPT and EDTA tubes. PPTs produce results equivalent to standard EDTA tube results when processed 4 h after phlebotomy. PPT intertube analyte results are reproducible. An average decrease of 13% and 37% in HIVL was observed in PPT plasma after 24 and 72 h of room temperature storage, respectively; thus, plasma can be stored at room temperature up to 24 h in the original tube. PPTs offer labor and supply savings over EDTA tubes.

Published 5 August 2005 in J Clin Microbiol, 43(8): 3769-71.
Full-text of this article is available online (may require subscription).

Place a permanent text-link or advertisement here for just US$15.

© 2004-2008 HIV Research Today. All Rights Reserved.



HIV Research Today Archive:

Volume 1 (2004)
  Issue 1 (August)
  Issue 2 (September)
  Issue 3 (October)
  Issue 4 (November)
  Issue 5 (December)

Volume 2 (2005)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 3 (2006)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 4 (2007)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)
  Issue 10 (October)
  Issue 11 (November)
  Issue 12 (December)

Volume 5 (2008)
  Issue 1 (January)
  Issue 2 (February)
  Issue 3 (March)
  Issue 4 (April)
  Issue 5 (May)
  Issue 6 (June)
  Issue 7 (July)
  Issue 8 (August)
  Issue 9 (September)



HIV Books

Blue Pills: A Positive Love Story

Blue Pills: A Positive Love Story